Analysis of T2 limitations and off‐resonance effects on spatial resolution and artifacts in echo‐planar imaging
367
Zitationen
3
Autoren
1990
Jahr
Abstract
Several aspects of blipped echo-planar imaging (EPI) are treated mathematically. An expression relating the necessary readout gradient strength and sampling time to the spatial resolution and readout duration is derived. It is shown how the net spatial resolution may be limited by the object's T2 characteristics and B0 field homogeneity, irrespective of the number of sampled points. Additionally, off-resonance effects result in a loss of spatial resolution and image distortion to a considerably greater degree than in conventional two-dimensional Fourier transform imaging. The extent of these effects is directly related to the time required to acquire the data matrix, and is therefore amplified when EPI is implemented on a standard commercial whole-body system which because of limited gradient performance uses necessarily longer sampling durations. Specific hardware modifications to a standard commercial imager are considered to allow successful EPI implementation. EPI image characteristics are compared quantitatively with those of conventional methods.
Ähnliche Arbeiten
Advances in functional and structural MR image analysis and implementation as FSL
2004 · 14.039 Zit.
A default mode of brain function
2001 · 12.343 Zit.
FSL
2011 · 11.670 Zit.
Improved Optimization for the Robust and Accurate Linear Registration and Motion Correction of Brain Images
2002 · 10.610 Zit.
Functional connectivity in the motor cortex of resting human brain using echo‐planar mri
1995 · 10.031 Zit.