MR fluoroscopy: Technical feasibility

1988·381 Zitationen·Magnetic Resonance in Medicine
Volltext beim Verlag öffnen

381

Zitationen

6

Autoren

1988

Jahr

Abstract

A method of magnetic resonance image acquisition and reconstruction is described in which high imaging rates and fast reconstruction times are allowed. The acquisition is a modification of the basic FLASH sequence but with a restricted number N of phase encodings. The encodings are applied sequentially, periodically, and continuously. Images are formed by sliding a window of width N encodings along the acquired data and reconstructing an image for each position of the window. In general the acquisition time per image exceeds the time between successive images, and the method thus has a temporal lag. Experimental studies were performed with a dynamic phantom using 48 phase encodings and a TR of 20 ms, for an image acquisition time of about 1 s. The image display rate in the reconstructed sequence was 12.5 images/s, and the image sequence portrayed the motion of the phantom. Additional studies were done with 24 encodings. It is shown how the sliding window technique lends itself to high-speed reconstruction, with each newly acquired echo used to quickly update the image on display. The combination of the acquisition technique described and a hardware implementation of the reconstruction algorithm can result in realtime MR image acquisition and reconstruction.

Ähnliche Arbeiten

Autoren

Institutionen

Themen

Advanced MRI Techniques and ApplicationsNMR spectroscopy and applicationsMedical Imaging Techniques and Applications
Volltext beim Verlag öffnen