A comparison of reconstruction algorithms for breast tomosynthesis
317
Zitationen
4
Autoren
2004
Jahr
Abstract
Three algorithms for breast tomosynthesis reconstruction were compared in this paper, including (1) a back‐projection (BP) algorithm (equivalent to the shift‐and‐add algorithm), (2) a Feldkamp filtered back‐projection (FBP) algorithm, and (3) an iterative Maximum Likelihood (ML) algorithm. Our breast tomosynthesis system acquires 11 low‐dose projections over a 50° angular range using an (CsI:Tl) flat‐panel detector. The detector was stationary during the acquisition. Quality metrics such as signal difference to noise ratio (SDNR) and artifact spread function (ASF) were used for quantitative evaluation of tomosynthesis reconstructions. The results of the quantitative evaluation were in good agreement with the results of the qualitative assessment. In patient imaging, the superimposed breast tissues observed in two‐dimensional (2D) mammograms were separated in tomosynthesis reconstructions by all three algorithms. It was shown in both phantom imaging and patient imaging that the BP algorithm provided the best SDNR for low‐contrast masses but the conspicuity of the feature details was limited by interplane artifacts; the FBP algorithm provided the highest edge sharpness for microcalcifications but the quality of masses was poor; the information of both the masses and the microcalcifications were well restored with balanced quality by the ML algorithm, superior to the results from the other two algorithms.
Ähnliche Arbeiten
New response evaluation criteria in solid tumours: Revised RECIST guideline (version 1.1)
2008 · 29.182 Zit.
Robust uncertainty principles: exact signal reconstruction from highly incomplete frequency information
2006 · 15.729 Zit.
Image processing with ImageJ
2004 · 11.903 Zit.
Fast robust automated brain extraction
2002 · 10.780 Zit.
Improved Optimization for the Robust and Accurate Linear Registration and Motion Correction of Brain Images
2002 · 10.612 Zit.