Virtual monochromatic imaging in dual‐source dual‐energy CT: Radiation dose and image quality
337
Zitationen
6
Autoren
2011
Jahr
Abstract
In dual-source dual-energy CT, optimal virtual monochromatic energy depends on patient size, dose partitioning, and the image quality metric optimized. With the optimal monochromatic energy, the noise level was similar to and the iodine CNR was better than that in 120 kV images for the same radiation dose. Compared to single-energy 80 kV images, the iodine CNR in virtual monochromatic images was lower for small to large phantom sizes.