Dies ist eine Übersichtsseite mit Metadaten zu dieser wissenschaftlichen Arbeit. Der vollständige Artikel ist beim Verlag verfügbar.
A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animals
257
Zitationen
3
Autoren
1999
Jahr
Abstract
A HIDAC-PET camera consisting essentially of 5 million 0.5 mm gas avalanching detectors has been constructed for small-animal imaging. The particular HIDAC advantage-a high 3D spatial resolution-has been improved to 0.95 mm fwhm and to 0.7 mm fwhm when reconstructing with 3D-OSEM methods incorporating resolution recovery. A depth-of-interaction resolution of 2.5 mm is implicit, due to the laminar construction. Scatter-corrected sensitivity, at 8.9 cps/kBq (i.e. 0.9%) from a central point source, or 7.2cps/kBq (543cps/kBq/cm/sup 3/) from a distributed (40 mm diameter, 60 mm long) source is now much higher than previous, and other, work. A field-of-view of 100 mm (adjustable to 200 mm) diameter by 210 mm axially permits whole-body imaging of small animals, containing typically 4 MBqs of activity, at 40 kcps of which 16% are random coincidences, with a typical scatter fraction of 44%. Throughout the field-of-view there are no positional distortions and relative quantitation is uniform to /spl plusmn/3.5%, but some variation of spatial resolution is found. The performance demonstrates that HIDAC technology is quite appropriate for small-animal PET cameras.
Ähnliche Arbeiten
New response evaluation criteria in solid tumours: Revised RECIST guideline (version 1.1)
2008 · 29.200 Zit.
Robust uncertainty principles: exact signal reconstruction from highly incomplete frequency information
2006 · 15.733 Zit.
Image processing with ImageJ
2004 · 11.903 Zit.
Fast robust automated brain extraction
2002 · 10.782 Zit.
Improved Optimization for the Robust and Accurate Linear Registration and Motion Correction of Brain Images
2002 · 10.616 Zit.