On cone‐beam computed tomography artifacts induced by titanium implants
282
Zitationen
3
Autoren
2009
Jahr
Abstract
Our theoretical and experimental results prove massive beam-hardening artifacts for a typical implant diameter and typical energies of up-to-date CBCT machines. Meaningful artifact reduction has to be based on more sophisticated mathematical modeling of the actual physical image acquisition process rather than on postprocessing of the erroneous results obtained from the rather crude reconstruction algorithms used presently.