Absolute light output of scintillators
424
Zitationen
5
Autoren
1997
Jahr
Abstract
The absolute light outputs of BGO, CsI(Tl) and some new Ce-doped crystals have been measured to an accuracy of about /spl plusmn/5% using calibrated XP2020Q photomultipliers and standard S3590.03 and S2740.03 photodiodes. The use of small crystals, 9 mm in diameter and 1 mm thick, reduces the corrections for imperfections in the light collection process and in the photoelectron collection by the photomultipliers. The measured light-output of 8500/spl plusmn/350 ph/MeV (photons/MeV) for the BGO crystals agree well with the earlier measurement done by Holl et al. (1988). The carried out study highlighted the importances of the spread in the published emission spectra of the crystals which seems to limit measurement accuracy. Finally, a simple comparative method of measuring light output to an accuracy of /spl plusmn/10%, is proposed using an uncalibrated XP2020 photomultiplier and a 1 mm thick BGO crystal as a standard.
Ähnliche Arbeiten
Mechanical metallurgy
1962 · 4.398 Zit.
Recent developments in Geant4
2016 · 3.949 Zit.
Algorithms for the rapid simulation of Rutherford backscattering spectra
1985 · 2.799 Zit.
Upconversion Nanoparticles: Design, Nanochemistry, and Applications in Theranostics
2014 · 2.606 Zit.
FastJet 2.4.1 user manual