Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents counts
539
Zitationen
3
Autoren
1990
Jahr
Abstract
True coincidence count (TCC) and noise equivalent count (NEC) curves were measured with a standardized 20-cm-diameter nylon cylinder for five different CTI/Siemens PET (positron emission tomography) scanners with several scanner-collimator combinations: (1) 831/08-12 with 1-mm collimator septa; (2) 933/08-12 and 933/08-16 with 3 to 1-mm tapered collimator septa; and (3) 931/08-12 with 3 to 1-mm tapered and a 1-mm collimator septa and the 931/08-16 with 3 to 1-mm tapered collimator septa. In addition, TCC and NEC curves on the 933/08-12 were compared with those from an Alderson brain phantom. In general, it is found that the TCC curves indicated peak count rates and activity levels that were as much as 50% higher than the corresponding values from NEC curves. The primary factor causing this difference is the noise effect of the randoms component. It is demonstrated that, compared to the Alderson brain phantom, the standard 20-cm cylinder is a poor predictor of count rate performance for PET brain imaging.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
Ähnliche Arbeiten
New response evaluation criteria in solid tumours: Revised RECIST guideline (version 1.1)
2008 · 29.134 Zit.
fastp: an ultra-fast all-in-one FASTQ preprocessor
2018 · 28.497 Zit.
Radiotherapy plus Concomitant and Adjuvant Temozolomide for Glioblastoma
2005 · 21.375 Zit.
Robust uncertainty principles: exact signal reconstruction from highly incomplete frequency information
2006 · 15.723 Zit.
Image processing with ImageJ
2004 · 11.903 Zit.