Performance evaluation of the microPET P4: a PET system dedicated to animal imaging
307
Zitationen
8
Autoren
2001
Jahr
Abstract
The microPET Primate 4-ring system (P4) is an animal PET tomograph with a 7.8 cm axial extent, a 19 cm diameter transaxial field of view (FOV) and a 22 cm animal port. The system is composed of 168 detector modules, each with an 8 x 8 array of 2.2 x 2.2 x 10 mm3 lutetium oxyorthosilicate crystals, arranged as 32 crystal rings 26 cm in diameter. The detector crystals are coupled to a Hamamatsu R5900-C8 PS-PMT via a 10 cm long optical fibre bundle. The detectors have a timing resolution of 3.2 ns, an average energy resolution of 26%, and an average intrinsic spatial resolution of 1.75 mm. The system operates in 3D mode without inter-plane septa, acquiring data in list mode. The reconstructed image spatial resolution ranges from 1.8 mm at the centre to 3 mm at 4 cm radial offset. The tomograph has a peak system sensitivity of 2.25% at the centre of the FOV with a 250-750 keV energy window. The noise equivalent count rate peaks at 100-290 kcps for representative object sizes. Images from two phantoms and three different types of laboratory animal demonstrate the advantage of the P4 system over the original prototype microPET. including its threefold improvement in sensitivity and a large axial FOV sufficient to image an entire mouse in a single bed position.
Ähnliche Arbeiten
New response evaluation criteria in solid tumours: Revised RECIST guideline (version 1.1)
2008 · 29.173 Zit.
Robust uncertainty principles: exact signal reconstruction from highly incomplete frequency information
2006 · 15.727 Zit.
Image processing with ImageJ
2004 · 11.903 Zit.
Fast robust automated brain extraction
2002 · 10.780 Zit.
Improved Optimization for the Robust and Accurate Linear Registration and Motion Correction of Brain Images
2002 · 10.611 Zit.