Measuring the rate of manual transcription error in outpatient point-of-care testing
54
Zitationen
2
Autoren
2018
Jahr
Abstract
Many point-of-care laboratory tests are manually entered into the electronic health record by ambulatory clinic staff, but the rate of manual transcription error for this testing is poorly characterized. Using a dataset arising from a duplicated workflow that created a set of paired interfaced and manually entered point-of-care glucose measurements, we found that 260 of 6930 (3.7%) manual entries were discrepant from their interfaced result. Thirty-seven of the 260 (14.2%) errors were discrepant by more than 20% and included potentially dangerous mistranscriptions. An additional 37 (14.2%) errors were due to inclusion of non-numeric characters. Staff-entered result flags deviated from the result flag generated in the laboratory information system in 5121 of 6930 (73.9%) pairs. These data demonstrate that clinically significant discrepancies for clinic-entered point of care results occurred at a rate of approximately 5 per 1000 results and they underline the importance of interfacing instruments when feasible.
Ähnliche Arbeiten
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
1986 · 47.397 Zit.
The meaning and use of the area under a receiver operating characteristic (ROC) curve.
1982 · 21.652 Zit.
Basic principles of ROC analysis
1978 · 6.051 Zit.
Methods for the determination of limit of detection and limit of quantitation of the analytical methods
2011 · 3.112 Zit.
All About Albumin: Biochemistry, Genetics, and Medical Applications
1995 · 3.103 Zit.