Low-Dose CT of the Abdomen: Evaluation of Image Improvement with Use of Noise Reduction Filters—Pilot Study
159
Zitationen
9
Autoren
2003
Jahr
Abstract
A prospective assessment of improvement in image quality at low-radiation-dose computed tomography (CT) of the abdomen by using noise reduction filters was performed. CT images acquired at standard and 50% reduced tube current were processed with six noise reduction filters and evaluated by three radiologists for image noise, sharpness, contrast, and overall image quality in terms of abdominal organ depiction. Quantitative image noise and contrast-to-noise ratio were measured. Baseline low-dose CT images were significantly worse than standard-dose CT images (P <.05). A statistically significant reduction of noise in low-dose images processed with three filters was noted. In conclusion, use of noise reduction filters decreased image noise at low-dose CT.