Towards dose reduction in CT: patient radiation dose assessment for CT examinations at university health center in Canada and comparison with national diagnostic reference levels

2011·28 Zitationen·Radiation Protection Dosimetry
Volltext beim Verlag öffnen

28

Zitationen

3

Autoren

2011

Jahr

Abstract

Dose indicators such as the computed tomography dose index (CTDI) and dose-length product (DLP) were gathered for all routine abdomen-pelvis, chest and head examinations performed on all computed tomography (CT) scanners at a University Health Center (UHC) in Canada. These indicators were analysed and compared with the range of diagnostic reference levels (DRLs) suggested by Health Canada and with DRLs in other countries. Mean DLP values varied from one scanner to another, but mean values at the UHC (750 mGy cm(-1) for abdomen-pelvis CT, 349 mGy cm(-1) for chest CT and 1181 mGy cm(-1) for head CT) were all below the upper limit of the range of DRLs suggested by Health Canada. Local DRLs at the UHC were set to 810 mGy cm(-1) for abdomen-pelvis CT, 345 mGy cm(-1) for chest CT and 1205 mGy cm(-1) for head CT. Results, however, show the need for protocols revisions, since some scanners exhibit mean DLP values slightly below or above the upper limit of the range of DRLs suggested by Health Canada.