Dies ist eine Übersichtsseite mit Metadaten zu dieser wissenschaftlichen Arbeit. Der vollständige Artikel ist beim Verlag verfügbar.
Patient doses in paediatric interventional cardiology: impact of 3D rotational angiography
28
Zitationen
4
Autoren
2015
Jahr
Abstract
The aim of this study was to calculate the contribution of 3D rotational angiography to radiation doses received by paediatric patients in a cardiac catheterisation laboratory. The percentage increase in the median value of air kerma-area product due to cone beam CT was 33 and 16% for diagnostic and therapeutic procedures, respectively. Results are presented separately for five age groups and ten weight groups. Several methods for reducing radiation from 3D rotational angiography are suggested and patient doses are compared with previously published values.
Ähnliche Arbeiten
Computed Tomography — An Increasing Source of Radiation Exposure
2007 · 8.640 Zit.
Maximum Likelihood Reconstruction for Emission Tomography
1982 · 4.367 Zit.
The 2007 recommendations of the International Commission on Radiological Protection
2007 · 4.110 Zit.
Sources and effects of ionizing radiation
2018 · 3.989 Zit.
Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study
2012 · 3.624 Zit.