Observer studies involving detection and localization: Modeling, analysis, and validation
349
Zitationen
2
Autoren
2004
Jahr
Abstract
Although the receiver operating characteristic (ROC) paradigm is the accepted method for evaluation of diagnostic imaging systems, it has some serious shortcomings inasmuch as it is restricted to one observer report per image. By contrast the free-response ROC (FROC) paradigm and associated analysis method allows the observer to report multiple abnormalities within each imaging study, and uses the location of reported abnormalities to improve the measurement. Because the ROC method cannot accommodate multiple responses or use location information, its statistical power will suffer. The FROC paradigm/analysis has not enjoyed widespread acceptance because of concern about whether responses made to the same diagnostic study can be treated as independent. We propose a new jackknife FROC analysis method (JAFROC) that does not make the independence assumption. The new analysis method combines elements of FROC and the Dorfman-Berbaum-Metz (DBM) methods. To compare JAFROC to an earlier free-response analysis method (specifically the alternative free-response, or AFROC method), and to the DBM method, which uses conventional ROC scoring, we developed a model for generating simulated FROC data. The simulation model is based on an eye-movement model of how experts evaluate images. It allowed us to examine null hypothesis (NH) behavior and statistical power of the different methods. We found that AFROC analysis did not pass the NH test, being unduly conservative. Both the JAFROC method and the DBM method passed the NH test, but JAFROC had more statistical power than the DBM method. The results of this comparison suggest that future studies of diagnostic performance may enjoy improved statistical power or reduced sample size requirements through the use of the JAFROC method.
Ähnliche Arbeiten
Refinement and reassessment of the SERVQUAL scale.
1991 · 3.967 Zit.
Features and uses of high-fidelity medical simulations that lead to effective learning: a BEME systematic review
2005 · 3.800 Zit.
Radiobiology for the Radiologist.
1974 · 3.502 Zit.
International evidence-based recommendations for point-of-care lung ultrasound
2012 · 2.832 Zit.
Radiation Dose Associated With Common Computed Tomography Examinations and the Associated Lifetime Attributable Risk of Cancer
2009 · 2.435 Zit.