Dies ist eine Übersichtsseite mit Metadaten zu dieser wissenschaftlichen Arbeit. Der vollständige Artikel ist beim Verlag verfügbar.
Female Breast Radiation Exposure During CT Pulmonary Angiography
269
Zitationen
6
Autoren
2005
Jahr
Abstract
CT pulmonary angiography delivers a minimum radiation dose of 2.0 rad (20 mGy) to the breasts of an average-sized woman. This greatly exceeds the American College of Radiology recommendation of < or = 0.300 rad (3 mGy) or less for standard two-view mammography. The potential latent carcinogenic effects of such radiation exposure at this time remain unknown. We encourage the judicious use of CT pulmonary angiography and lower doses and nonionizing radiation alternatives when appropriate.