Individually Adapted Examination Protocols for Reduction of Radiation Exposure in Chest CT

2001·74 Zitationen·Investigative Radiology
Volltext beim Verlag öffnen

74

Zitationen

8

Autoren

2001

Jahr

Abstract

By deriving mAs values from body weight estimation, an individually adapted protocol for chest CT can be recommended and easily employed in a clinical setting. With an adaptation of the tube current-time product based on the estimated body weight of the patient - 10 (body weight [kg] - 10 mAs), a well-balanced examination without significant loss of information, even in soft-tissue window settings, can be performed with this particular scanner. For this adapted mAs protocol, a mean reduction of radiation exposure of 45% was achievable, compared with the standard protocol. A maximum decrease per case down to 31 mAs was obtained, without relevant loss of image quality. Therefore, for other types of CT scanners, analogous protocols may be adapted.